Brandtech-scientific QuikSip BT-Aspirator Instrukcja Użytkownika Strona 18

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 17
- 18 -
Troubleshooting
Problem Possible cause Corrective action
Will not aspirate
Dispensing cartridge not locked
Dispensing cartridge leaks
Filling valve leaks or is sticking
Connection between filling tube and
tube of the cell-culture
leaks
Tube of cell-culture
blocked or sticking
Lock dispensing cartridge ( page 8).
Mount new dispensing cartridge ( page 13).
Clean filling valve ( page 16) or replace valve and seal.
Stick adapter firmly onto the filling tube and pull tube firmly
onto the adapter.
Clean tube or replace.
Piston and dispensing
cartridge difficult to move
Crystal deposition
Improper liquid aspirated
Replace dispensing cartridge ( page 13).
Check, if liquid is proper to aspirate.
Liquid behind the piston of
the dispensing cartridge
Dispensing cartridge leaks
Replace dispensing cartridge ( page 13).
Aspirated liquid can't be
delivered into the bottle
Discharge valve sticking or damaged
Clean valve (page 16) or replace.
Przeglądanie stron 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 27 28

Komentarze do niniejszej Instrukcji

Brak uwag