Brandtech-scientific EASYCAL 4.0 Instrukcja Użytkownika Strona 1

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Urządzenia pomiarowe Brandtech-scientific EASYCAL 4.0. BrandTech Scientific EASYCAL 4.0 User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 36
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 0
EASYCAL
User Guide
Version 4.0
Przeglądanie stron 0
1 2 3 4 5 6 ... 35 36

Podsumowanie treści

Strona 1 - User Guide

EASYCAL™ User GuideVersion 4.0

Strona 2

10 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User GuideNote:The primary data according to GLP instructions will be transmitted wit

Strona 3

EASYCAL™ User Guide Version 4.0 · Status: June 2008 11EASYCAL™ User GuideIn case the instrument is tested for the first time you will be asked at whi

Strona 4

12 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User Guide3.2 Calibration of volumetric instruments of glass/plasticClick on the gra

Strona 5

EASYCAL™ User Guide Version 4.0 · Status: June 2008 13EASYCAL™ User GuideSubdivision (required entry)Click on the arrow to the right of the entry fie

Strona 6

14 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User GuideEntry of measuring results☞ The entry fields for measuring results have a

Strona 7

EASYCAL™ User Guide Version 4.0 · Status: June 2008 15EASYCAL™ User Guide3.2.2 Entry of error limitsOn the “Entry of tolerance limits” screen you de

Strona 8

16 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User Guide3.2.4 “Save present test” (permanently) and “Intervall entry”4.1 Selectin

Strona 9

EASYCAL™ User Guide Version 4.0 · Status: June 2008 17EASYCAL™ User GuideDefining the filter (Example):Example:☞ Use the mouse to select an option f

Strona 10

18 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User Guide4.7 GraphThe “Curve” button allows you to view a graph of the variouscalibr

Strona 11

EASYCAL™ User Guide Version 4.0 · Status: June 2008 19EASYCAL™ User Guide6.1 Company and LanguageClick on “Please define company and language” to di

Strona 13

20 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User GuideCreate a new test recordClick “New” or “F2” to open a blank entry screen. E

Strona 14

EASYCAL™ User Guide Version 4.0 · Status: June 2008 21EASYCAL™ User Guide6.4 Manage EASYCAL™ dataClick on “Manage Easycal data” to open the following

Strona 15

22 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User Guide“Read test archives”Data which have been saved as an archive can be re-impo

Strona 16

EASYCAL™ User Guide Version 4.0 · Status: June 2008 23EASYCAL™ User GuideChange PasswordYou can change the preferences and the password to protect yo

Strona 17

24 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User GuideCoeffi cient of VariationThe coefficient of variation (CV) is a measure of r

Strona 18

EASYCAL™ User Guide Version 4.0 · Status: June 2008 25EASYCAL™ User GuidePoor accuracy:Hits far off center.Poor reproducibility:Hits widely scattered

Strona 19

26 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User Guide7.2 BasicsThe values obtained by weighing during the gravimetric test are o

Strona 20

EASYCAL™ User Guide Version 4.0 · Status: June 2008 27EASYCAL™ User Guide7.3 Calculations for Glass/PlasticsExample:Serial No: 95 080634Trademark:

Strona 21

28 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User GuideDefinitions in Test Record for Glass/PlasticsDeviation = Maximum deviation

Strona 22

EASYCAL™ User Guide Version 4.0 · Status: June 2008 29EASYCAL™ User GuideTable 1Table 1 provides the factor “Z” for temperatures from 15 °C to 30 °C

Strona 23

EASYCAL™ User Guide Version 4.0 · Status: June 2008 3EASYCAL™ User GuideContents1. Your License Agreement ...

Strona 24

30 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User GuideTable 2If volumetric instruments made of plastic are to be tested, the foll

Strona 25

EASYCAL™ User Guide Version 4.0 · Status: June 2008 31EASYCAL™ User Guide7.4 Calculations for Liquid HandlingExcerpt from ISO 8655/3, Table refers t

Strona 26

32 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User Guide8.1 Printing problemsParts of the record are not printedThis problem occurs

Strona 27

EASYCAL™ User Guide Version 4.0 · Status: June 2008 33EASYCAL™ User Guide8.2 Memory problemsThe MS ACCESS database application, which forms the basis

Strona 28

34 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User GuideAccuracy ... 24Adopt

Strona 30

BRAND GMBH + CO KG · P.O. Box 1155 · 97861 Wertheim · GermanyPhone: +49 9342 808-0 · Fax: +49 9342 808-236 · E-Mail: [email protected]: www.brand.

Strona 31

4 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User Guide1. Your Licence Agreement1.1 Object of Agreement The object of this agree

Strona 32

EASYCAL™ User Guide Version 4.0 · Status: June 2008 5EASYCAL™ User Guide2.1 Program featuresEASYCAL™ is a software system that was developed using Mi

Strona 33

6 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User Guide2.4 Uninstalling the programPrograms that have been specially developed for

Strona 34

EASYCAL™ User Guide Version 4.0 · Status: June 2008 7EASYCAL™ User Guide3. Main Menu3.1 Calibration of Liquid Handling InstrumentsClick on the gray

Strona 35

8 EASYCAL™ User Guide Version 4.0 · Status: June 2008 EASYCAL™ User GuideEnter the number or the type of the thermometer used in the test. By clickin

Strona 36 - Internet: www.brand.de

EASYCAL™ User Guide Version 4.0 · Status: June 2008 9EASYCAL™ User GuideTest temperature (required entry)The permissible temperature range for measur

Komentarze do niniejszej Instrukcji

Brak uwag